3500xL Dx Genetic Analyzer 24-Capillary Array, 50 cm
These internally uncoated 50 cm capillaries are supplied in preassembled sets assemblies of 24 capillaries per array, with a built-inRead more
Have Questions?
Catalog Number
For Use With (Application)
For Use With (Equipment)
4404688
Sequencing
3500xL Dx Genetic Analyzer, 3500 Genetic Analyzer
Catalog number 4404688
Price (USD)
3,240.00
Each
Add to cart
For Use With (Application):
Sequencing
For Use With (Equipment):
3500xL Dx Genetic Analyzer, 3500 Genetic Analyzer
Price (USD)
3,240.00
Each
Add to cart
These internally uncoated 50 cm capillaries are supplied in preassembled sets assemblies of 24 capillaries per array, with a built-in frame for easy installation. Radio frequency identification (RFID) tag is built into the frame for tracking critical run information, which is then reported on the software dashboard. The arrays may be used for 160 injections in Dx mode.
This product is intended for in-vitro diagnostic use on the Applied Biosystems 3500 Dx Series Genetic Analyzers and is not validated for use with other systems or applications. Refer to the Applied Biosystems 3500 Dx⁄3500xL Dx Genetic Analyzer User Guide for further details.
For in-vitro diagnostic use
Specifications
Capillary Length50 cm
For Use With (Equipment)3500xL Dx Genetic Analyzer, 3500 Genetic Analyzer
For Use With (Application)Sequencing
Product Type3500xL Dx Genetic Analyzer
Quantity1 array
Shipping ConditionRoom Temperature
Unit SizeEach
Contents & Storage
Ambient, 1 array with 24 capillaries
Frequently asked questions (FAQs)
When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?
Some of the other causes of an Unstable Electrophoresis current detected error message are:
1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues
Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.
Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?
If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.